Functional
Diversified Reciprocating Mode
Imported Sliding Table with Original Packaging
Single-Vessel, Independent,Fully-Automatic Evaporation Cover
500mL Large-Volume Liquid Measuring Cup
Optional Dual-Row 14 Positions
Optional Imaging Module
Safety Compliance
Compliant with CHP, USP, EP, and BP technical requirements.
Three levels of the management system and operation log in compliant with FDA 21 CFR Part 11 requirements to ensure the validity and reliability of electronic data.
The interface to LIMS laboratory information managementsystem for easy connection.
Multiple alarm modes, including sampling time and temperature overload alarms, are available.
Accuracy and Reliability
Andisso USP Apparatus 3 has mechanical calibration service.
The original imported slide table is used for stable and reliable operation, and the displacement repeatability can reach ±0.025mm under unloaded conditions.
The dissolution cup is equipped with an independently automatic anti-evaporation lid design, which can reduce the evaporation of the dissolution medium while effectively preventing cross-contamination.
Extensive Functions
4-60dpm reciprocation frequency meets the needs of both conventional and high-frequency reciprocation.
The rack automatically returns to the front row for further useafter a cycle.
Diversified reciprocation modes are available. When sampling multiple points in a single row, the reciprocation frequency canbe changed at different points in time.
The 500mL dissolution cup supports the multiple times of same temperature replenishment during the experiments.
An independent cleaning path.
Pre-heating function and high-power heating are available to quickly reach the set temperature.
Optional 14 positions in double rows for simultaneous testing of two groups of samples to improve experimental efficiency.
The camera module is optional to clearly record the experimental process and machine operation status.
Technical Specifications
USP Apparatus 3 | Autosampler |
---|---|
Instrument specifications: 7 positions and 6 rows | Sampling pipeline: 7-position syringe pump sampling |
Evaporation volume: less than 1% in 24 hours at 37℃ under static condition | Sample collection: 20 rows x 12 positions |
Temperature control range: room temperature to 50℃ | Test tube / liquid phase vials: 240 positions |
Dissolution vessel volume: 300mL | Sampling range: 0.5mL ~ 20mL |
Temperature control accuracy: ±0.2℃ | Maximum sampling times: 20 times |
Movement of reciprocating frame: 1 ~ 10cm | Continuous working time: 3333 hours |
Reciprocation frequency: 4 ~ 60dpm | Working noise: ≤50dB |