ChP
Flow-through cell
Standard inner diameter of a flow-through cell:12mm/22.6mm
The cone is filled with glass beads of 1.0mm
diameter, and a glass bead of 5.0mm diameter is placed at the bottom of the inverted cone to prevent the medium in the cell from flowing back into the pipeline.
Material: made of transparent inert material
Pump
The dissolution medium flows upwards through the flow-through cell under the action of the pump.
Commonly used flow rate: 240mL/h-960mL/h
Standard flow rate: 4mL/min, 8mL/min and 16mL/min
Flow rate curve: sinusoid
Pulsation frequency: 120 pulses/min ± 10 pulses/min